模擬実験シーケンス : |
#100700 ~ #100701 |
プラズマ実験 : |
#100702 ~ #100793 合計 92 ショット | 稼働率 : | 99.76 % | 特記事項 : | |
|
収集計測 : 合計 89 計測 ※グレー表示は一部収集
|
ARMS | AXUVD | AXUVD-PXI | Bolometer | Brems | CDP | CNPA | CTRLDISP1 | CTRLDISP2 | CTRLDISP3 | Divertor-Interferometer | DivTC | DTS | EB-NPA | ECEI1 | ECEI2 | ECH | ECHdiag | Fastion | FIG | FIR-PXI | FIR-WE7000 | FIR1 | FIR2 | Flxloop | FMD | FMD2 | FPellet | GASPUFF | GPCRADH | Halpha | HDLP | HIBP | HIBPcp | HICO2 | ICHPXI | ICHVOLT | ImpMon | Langmuir | Langmuir2 | Langmuir3 | Langmuir4 | Langmuir5 | LiBP | LIP-ICCD | LIP-PMT-PXI | LostIon | Magnetics | Magnetics2 | Magnetics3lab1 | Magnetics3lab2 | MIC | MIR | MIR2 | MIR3 | MIR4 | MMimg | MMWINT | MWRM | MWRM-PXI | MWRM-PXI2 | mwscat | NB1arm | NB2arm | NB3arm | NB4arm | NB5arm | NBEL | NBEL-RT | P2CO2 | PCO2 | PelletAblation | PHA | PHA-WE7000 | PHARD | RADH | RADL | RADLWE | Reflectmetry | RePellet | SDNPA | SiFNA | SiFNA-WE | SX8O | SXfluc | SXmp | TESPEL | TESPEL-CCD | TSCCD |
|
休止計測 : |
CCDedge1 | Demodulator | DivIR | IRcamera-edge |
|
共通コメント : |
No log |
サイトコメント : |
計測名/ショット№/モード |
コメント |
更新日時/書込者 |
| 起動確認 OK 点検(9:05) OK | 2010-11-18 09:07:24 今津 |
|
| 起動確認 OK 点検(9:05) OK(AMP slot#11,12,13 継続 slot#2 正常) | 2010-11-18 09:07:59 今津 |
|
| 起動確認 OK 点検(9:05) OK | 2010-11-18 09:07:24 今津 |
|
| キャンセルショット→#100785, #100788, #100792 ,#100793(真空度悪化のため中断、本日の実験はこのショットで終了) | 2010-11-19 09:04:28 小嶋 |
|
| #100786-100788は、繰り返し用デモジュレータの動作確認のため、繰り返し動作なし | 2010-11-18 17:34:41 小嶋 |
|
|
収集中断 : 障害 : 19 ショット 停止 : 0 ショット |
中断数 |
理由 |
計測名/ショット№ |
エラー内容/対処 |
更新日時/書込者 |
7 | 障害 | FIR-WE7000 #100702 ~ #100708 | WE通信エラーのため収集できず 光ファイバ接続の確認、WEシャーシの電源再投入などを何度か行って復旧 | 2010-11-18 10:01:01 大砂 |
|
1 | 障害 | LiBP #100750 | Ch.25-28 NotCompleted
| 2010-11-18 12:35:45 大砂 |
|
3 | 障害 | Langmuir #100784 ~ #100786 | 9分22秒周期へ切り替え後、外部クロック信号が正しく接続されていなかったためNot Completed(ch.13以外)
| 2010-11-18 17:24:44 小嶋 |
|
8 | 障害 | MWRM #100784 ~ #100791 | Trig.Ch.3のスタートチャネルが繰り返し用のch.2に変更されていなかった
| 2010-11-18 17:45:18 小嶋 |
|
|